features

OXYGEN ANALYZER OXITEC5000  
The gas sealed cell is produced by a special sealing technology!
High precision, high stability, and long life duration are ensured!
  • Automatic startup
  • Packaged with installation accessories and cable
  • Functions are kept by the minimum maintenance.
  • Components can be replaced by the customer.
  • Gas sampling and processing are not needed.
  • Suitable for application under high temperature conditions (max. 1400°C)

operating principles

The oxygen concentration in a gas is measured from the difference in the oxygen partial pressure (pressure that oxygen accounts for in the atmospheric pressure) between the reference gas in the meter body and the measured gas.  

standard specification

Items OXITEC 5000
Function Sensitivity   Below 1ppm O2
Accuracy   ±0.2%(from measured value)
Resolution
(A/D Converter)
  14 Bit
Controller Housing   A)Wall-hanging Type with a window box SS,IP-65
    400mmx300mmx225mm
  B)Built-in Type with a window box SS,IP-54
    177mmx483mmx360mm
  C)Wall-hanging explosion-proof Type
    with a round window box AC
    280mmx300mmx240mm
Power Voltage   115VAC or 230VAC 50/60Hz
Power Consumption   100-200VA(400VA at start-up)
Recommended Fuse   10A
Temperature   -20~+55℃(other temp on request)
Range   2 measuring ranges,0-99% adjustable
Output Analogue   0/4-20mA max.500Ω resistance load
Indication   System error,maintenance
Relay contact   System error,maintenance
  Upper limit,Lower limit
  Contact capacity : 230VAC 5A Resistance load
Display   LCD/LED Graphic display
Interface   RS232C,RS485,bus compatible

  • product lineup

    • OXITEC5000

    • OXITEC ECONOMY